Cross-Calibration and Comparison of Variability in 2 Bone Densitometers in a Research Setting: The Framingham Experience

  • David R. Gagnon

      Affiliations

    • Department of Biostatistics, Boston University School of Public Health, Boston, MA, USA
    • VA Cooperative Studies Program, Massachusetts Veterans Epidemiology Research and Information Center (MAVERIC), Boston, MA, USA
  • ,
  • Robert R. McLean

      Affiliations

    • Institute for Aging Research, Hebrew SeniorLife, Boston, MA, USA
    • Harvard Medical School, Boston, MA, USA
  • ,
  • Marian T. Hannan

      Affiliations

    • Institute for Aging Research, Hebrew SeniorLife, Boston, MA, USA
    • Harvard Medical School, Boston, MA, USA
    • Department of Epidemiology, Boston University School of Public Health, Boston, MA, USA
  • ,
  • L. Adrienne Cupples

      Affiliations

    • Department of Biostatistics, Boston University School of Public Health, Boston, MA, USA
  • ,
  • Mary Hogan

      Affiliations

    • Institute for Aging Research, Hebrew SeniorLife, Boston, MA, USA
  • ,
  • Douglas P. Kiel

      Affiliations

    • Institute for Aging Research, Hebrew SeniorLife, Boston, MA, USA
    • Harvard Medical School, Boston, MA, USA
    • Corresponding Author InformationAddress correspondence to: Douglas P. Kiel, MD, MPH, Institute for Aging Research, Hebrew SeniorLife, 1200 Centre Street, Boston, MA 02131.

Received 15 December 2008; received in revised form 30 December 2009; accepted 1 January 2010. published online 29 March 2010.

Abstract 

New technology introduced over time results in changes in densitometers during longitudinal studies of bone mineral density (BMD). This requires that a cross-calibration process be completed to translate measurements from the old densitometer to the new one. Previously described cross-calibration methods for research settings have collected single measures on each densitometer and used linear regression to estimate cross-calibration corrections. Thus, these methods may produce corrections that have limited precision and underestimate the variability in converted BMD values. Furthermore, most of the previous studies have included small samples recruited from specialized populations. Increasing the sample size, obtaining multiple measures on each machine, and using linear mixed models to account for between- and within-subject variability may improve cross-calibration estimates. The purpose of this study was to conduct an in vivo cross-calibration of a Lunar DPX-L (Lunar Corporation, Madison, WI) with a Lunar Prodigy densitometer (GE Medical Systems Lunar, Madison, WI) using a sample of 249 healthy volunteers who were scanned twice on each densitometer, without repositioning, at both the femur and spine. Scans were analyzed using both automated and manual placement of regions of interest. Wilcoxon rank-sum tests and Bland-Altman plots were used to examine possible differences between repeat scans within and across densitometers. We used linear mixed models to determine the cross-calibration equations for the femoral neck, trochanter, total hip, and lumbar spine (L2–L4) regions. Results using automated and manual placement of the regions of interest did not differ significantly. The DPX-L densitometer exhibited larger median absolute differences in the BMD values by repeat scans of femoral neck (0.016 vs 0.012, p=0.1) and trochanter (0.011 vs. 0.009, p=0.06) compared with the Prodigy densitometer. The Bland-Altman plots revealed no statistically significant linear relationship between the differences in paired measures between machines and mean BMD. In our large sample of healthy volunteers, we did detect systematic differences between the DPX-L and Prodigy densitometers. Our proposed cross-calibration method, which includes acquiring multiple measures and using linear mixed models, provides researchers with a more realistic estimate of the variance of cross-calibrated BMD measures, potentially reducing the chance of making a type I error in longitudinal studies of changes in BMD.

Key Words: Bone, mineral density, cross-calibration, dual x-ray absorptiometry

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1094-6950(10)00004-1

doi:10.1016/j.jocd.2010.01.003

Journal of Clinical Densitometry
Volume 13, Issue 2 , Pages 210-218, April 2010