Journal of Clinical Densitometry
Volume 11, Issue 2 , Pages 276-282 , April 2008

In Vivo and In Vitro Comparison of Densitometers in the NOREPOS Study

  • Tone K. Omsland

      Affiliations

    • Institute of General Practice and Community Medicine, University of Oslo, Oslo, Norway
    • Corresponding Author InformationAddress correspondence to: Tone Kristin Omsland, Section for Preventive Medicine and Epidemiology, Institute of General Practice and Community Medicine, Faculty of Medicine, University of Oslo, Oslo, Norway.
  • ,
  • Nina Emaus

      Affiliations

    • Institute of Community Medicine, University of Tromsø, Tromsø, Norway
  • ,
  • Clara G. Gjesdal

      Affiliations

    • Department of Public Health and Primary Health Care, University of Bergen, Bergen, Norway
  • ,
  • Jan A. Falch

      Affiliations

    • Center of Endocrinology, Aker University Hospital and Faculty of Medicine, Oslo, Norway
  • ,
  • Grethe S. Tell

      Affiliations

    • Department of Public Health and Primary Health Care, University of Bergen, Bergen, Norway
  • ,
  • Lisa Forsen

      Affiliations

    • Division of Epidemiology, Norwegian Institute of Public Health, Oslo, Norway
  • ,
  • Gro K.R. Berntsen

      Affiliations

    • Institute of Community Medicine, University of Tromsø, Tromsø, Norway
  • ,
  • Haakon E. Meyer

      Affiliations

    • Institute of General Practice and Community Medicine, University of Oslo, Oslo, Norway
    • Division of Epidemiology, Norwegian Institute of Public Health, Oslo, Norway

Received 21 June 2007 ,Revised 4 October 2007 ,Accepted 5 October 2007.

References 

  1. Blake GM, Fogelman I. Technical principles of dual energy x-ray absorptiometry. Semin Nucl Med. 1997;27:210–228
  2. Kolta S, Ravaud P, Fechtenbaum J, Dougados M, Roux C. Accuracy and precision of 62 bone densitometers using a European Spine Phantom. Osteoporos Int. 1999;10:14–19
  3. Pearson D, Cawte SA, Green DJ. A comparison of phantoms for cross-calibration of lumbar spine DXA. Osteoporos Int. 2002;13:948–954
  4. Genant HK, Grampp S, Glüer CC, et al. Universal standardization for dual x-ray absorptiometry: patient and phantom cross-calibration results. J Bone Miner Res. 1994;9:1503–1514
  5. Pearson J, Dequeker J, Henley M, et al. European semi-anthropomorphic spine phantom for the calibration of bone densitometers: assessment of precision, stability and accuracy. The European Quantitation of Osteoporosis Study Group. Osteoporos Int. 1995;5:174–184
  6. Kalender WA, Felsenberg D, Genant HK, Fischer M, Dequeker J, Reeve J. The European Spine Phantom—a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT. Eur J Radiol. 1995;20:83–92
  7. Oldroyd B, Smith AH, Truscott JG. Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers—bone mineral density and body composition studies. Eur J Clin Nutr. 2003;57:977–987
  8. Cawte SA, Pearson D, Green DJ, Maslanka WB, Miller CG, Rogers AT. Cross-calibration, precision and patient dose measurements in preparation for clinical trials using dual energy X-ray absorptiometry of the lumbar spine. Br J Radiol. 1999;72:354–362
  9. Blake GM, Harrison EJ, Adams JE. Dual X-ray absorptiometry: cross-calibration of a new fan-beam system. Calcif Tissue Int. 2004;75:7–14
  10. Blake GM. Replacing DXA scanners: cross-calibration with phantoms may be misleading. Calcif Tissue Int. 1996;59:1–5
  11. Steiger P. Standardization of postero-anterior (PA) spine BMD measurements by DXA. Committee for Standards in DXA. Bone. 1995;17:435
  12. Pearson D. Cross calibration of DXA as part of an equipment replacement program. J Clin Densitom. 2006;9:287–294
  13. Mazess RB, Hanson JA, Payne R, Nord R, Wilson M. Axial and total-body bone densitometry using a narrow-angle fan-beam. Osteoporos Int. 2000;11:158–166
  14. Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK. Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int. 1995;5:262–270
  15. Bland JM, Altman DG. Measurement error. BMJ. 1996;313:744
  16. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–160
  17. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310
  18. Fitzmaurice GM. Applied longitudinal analysis. New Jersey: John Wiley & Sons; 2004;
  19. Saarelainen J, Honkanen R, Vanninen E, et al. Cross-calibration of Lunar DPX-IQ and DPX dual-energy x-ray densitometers for bone mineral measurements in women: effect of body anthropometry. J Clin Densitom. 2005;8:320–329
  20. Mazess RB, Barden HS. Evaluation of differences between fan-beam and pencil-beam densitometers. Calcif Tissue Int. 2000;67:291–296
  21. Crabtree NJ, Shaw NJ, Boivin CM, Oldroyd B, Truscott JG. Pediatric in vivo cross-calibration between the GE Lunar Prodigy and DPX-L bone densitometers. Osteoporos Int. 2005;16:2157–2167

 The NOREPOS Research Group: The Core Research Group (in alphabetic order): Nina Emaus (Tromsø), Haakon E. Meyer (Oslo), Berit Schei (Nord-Trøndelag), and Grethe S. Tell (Bergen). Local collaborators (in alphabetic order): Bergen: Gjesdal CG, Tell GS; Nord-Trøndelag: Forsén L, Forsmo S, Langhammer A, Schei B; Oslo: Falch JA, Meyer HE, Søgaard AJ; Tromsø: Berntsen GKR, Emaus N, Fønnebø V, Joachimsen RM.

PII: S1094-6950(07)00211-9

doi: 10.1016/j.jocd.2007.10.001

Journal of Clinical Densitometry
Volume 11, Issue 2 , Pages 276-282 , April 2008