Journal of Clinical Densitometry
Volume 10, Issue 3 , Pages 249-258, July 2007

A Generalized Least Significant Change for Individuals Measured on Different DXA Systems

  • John A. Shepherd

      Affiliations

    • Corresponding Author InformationAddress correspondence to: John A. Shepherd, PhD, Department of Radiology, University of California, 185 Berry Street, Suite 350, San Francisco, CA 94143-0946.
  • ,
  • Ying Lu

Department of Radiology, University of California, San Francisco, CA

Received 28 April 2006; received in revised form 5 May 2007; accepted 7 May 2007. published online 09 July 2007.

Abstract 

In this article, we derive a generalized expression for the least significant change (LSC), which we call the generalized LSC (GLSC), to be used when an individual is measured on 2 different systems. The commonly used LSC is defined as the least amount of change between 2 measurements over time that must be exceeded before a change can be considered true (with 95% confidence). The LSC has clinical applications in monitoring disease progression or treatment effects in bone mineral density (BMD) and bone mineral content. Mathematically, the “ideal” LSC (ILSC) is 2.77 times the precision errors for measures on a single system. When BMD values of an individual are measured by 2 different systems, the LSC will depend not only on the precision errors of both systems but also on the calibration relationship between the systems. Like the ILSC, the GLSC is a simple equation applicable for inter machine comparisons. The GLSC can be defined for any 2 systems with measures obtained from cross-calibration and precision studies using the protocols recommended by the International Society for Clinical Densitometry. We validated the GLSC using 10,000 simulated measurements taken between 2 systems and offer several common uses of the GLSC such as system upgrades within a single manufacturer and replacement of 1 manufacturer by another. We found that when upgrading a Hologic QDR-2000 to a QDR-4500, GLSC was twice as large as the QDR-2000 LSC (0.0432 and 0.0215g/cm2, respectively). The GLSC was 2.6 (spine) to 3.6 (total hip) times larger than the LSC when comparing scans between the Hologic Delphi and the GE Lunar Prodigy. We also explore how the magnitude of the correlation coefficient and sample size change the GLSC and show that a correlation coefficient less than 0.95 increases the %GLSC to above 10%, and that increasing study sample sizes beyond 30 in the cross-calibration studies can only decrease the magnitude of the GLSC accuracy by 4%. We conclude that the GLSC, defined using commonly used clinical cross-calibration and precision assessments, is the most accurate method to compare scans between dual-energy X-ray absorptiometry systems.

Key Words: Accuracy, bone mineral density (BMD), cross-calibration, dual-energy X-ray absorptiometry (DXA), GE healthcare, hologic, least significant change (LSC), precision

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1094-6950(07)00173-4

doi:10.1016/j.jocd.2007.05.002

Journal of Clinical Densitometry
Volume 10, Issue 3 , Pages 249-258, July 2007